THIS SERVICE PACK IS FOR METEX 1.2 USERS WITH MB VERSION 7.1X ONLY!!!

INSTALLATION INSTRUCTIONS FOR METEX_123_SP3.ZIP  service pack 3

1 - Unzip this archive on the administrators pc where the Metrology Administration program is located.
    Usually by default in the local "C:\METCAL" folder.

2 - Copy the MET_EX.EXE file to the server in the ......\cgi-bin folder.

3 - Copy the ".css" style sheets files to the cgi-bin\htdocs directory.

4 - Reregister METEX 1.2 from the folder that you unzipped the archive to.


CHANGES BY THIS SERVICE PACK INCLUDES SP1 and SP2

1 - Some Crystal Reports with prompting formula fields did not get any data or got a mismatch error.

2 - Users that have multiple desktops would see in the Calibration form web page multiple Standards Used - C2340 fields depending on the number of desktops.

3 - Sorting of cal events for selection when viewing or printing a Crystal Cal Certificate only sorting by date. Should also sort by time of day to catch the AS-FOUND and AS-LEFT cal events in the right order.

4 - Tweeked the METEX Administration program to keep the offset for field C2340.

5 - Fixed error "9, Subscript out of range". This error occurred when using more than 3 tables.

6 - Fixed a problem when a text field such as Model 773-10 was getting converted to a date "10/01/0733".

7 - Added additional lines in style sheets to select font-size and type so that the MT forms where getting all the data without truncation.

8 - Changed a stored procedure to allow the hiding of the Standards Used Field per Group ID.

9 - Changed the default session start date and stop date to todays date instead of April 1999.

10 - Changed registration scripts to grant select persmission to tables even if they already exist.

11. Chnaged the Description Field in the full result form size from 25 characters to 32 characters.